Microbiological Safety Cabinet Testing & Servicing

Testing and servicing of Class I and Class II microbiological safety cabinets to BS EN 12469, including airflow verification, HEPA integrity, and KI Discus containment testing.

Home > Services > Microbiological Safety Cabinet Testing & Servicing

Why Safety Cabinets Need Specialist Testing

Microbiological safety cabinets protect operators, the product, and the environment from hazardous biological agents. They rely on precise airflow patterns and HEPA filtration to maintain containment. Any deterioration in performance, whether from filter loading, fan degradation, or seal failure, compromises that protection.

Regular testing to BS EN 12469 confirms the cabinet is still providing the containment it was designed for.

What We Test

  • Downflow and inflow velocity measurement
  • Airflow uniformity across the work opening
  • HEPA filter integrity testing (DOP/DPC)
  • KI Discus operator protection test
  • Cabinet integrity (pressure hold test)
  • Alarm function verification
  • UV lamp intensity (where fitted)
  • Noise levels

Servicing

  • HEPA filter replacement (supply and exhaust)
  • Pre-filter changes
  • Fan motor servicing
  • UV lamp replacement
  • Airflow calibration
  • Decontamination (formaldehyde or hydrogen peroxide) prior to filter changes or relocation

Cabinet Types We Service

  • Class I (operator protection)
  • Class II Type A (operator and product protection)
  • Class II Type B (operator, product, and environment protection)
  • Laminar flow cabinets and clean air workstations

Frequently Asked Questions

How often should safety cabinets be tested?

At least annually, after any filter change, after relocation, and after any maintenance that may affect airflow. Many containment facilities test every 6 months.

What is KI Discus testing?

KI Discus is a containment verification test that challenges the cabinet’s operator protection by generating a potassium iodide aerosol inside the cabinet and sampling outside. It confirms the cabinet is containing biological hazards.

Do you decontaminate before servicing?

Yes. Cabinets handling hazardous biological agents must be decontaminated before any internal work including filter changes. We carry out formaldehyde or hydrogen peroxide decontamination as appropriate.